Kestrel® RH Calibration Kit Each
Kestrel Pocket weather Meter is individually calibrated after
assembly. In essence, each unit is "taught" what the correct relative humidity
(RH) values should be for given signals from the RH sensor. The response of the humidity sensor is a highly reproducible, non-linear curve. This non-linear characteristic is stored in the microprocessor used to compute RH, and in this manner we are able to maintain an accuracy of ± 3% of scale over the relative range 5% - 95%. Field Calibration The Kestrel should be recalibrated if the RH sensor has been replaced or has shown drift in its RH measurement. Field calibration requires the use of the RH Calibration Kit. For instructions related to the field calibration procedure, please view the Field Calibration Kit Manual. This manual explains in detail all of the steps for recalibrating your Kestrel. This process will take approximately 13 hours, including setup and calibration. Contents of the RH Calibration Kit The RH calibration kit includes the following items:
The large container is used to create a stable environment. The small containers are used to hold the saturated salt solution. The saturated salt solutions generate well-defined humidities when they are confined to a contained environment.
|
|||